Applying High Voltage to a Segmented (Clover) HPGe Detector
- With
power off, connect the cables in the following order:
- Preamp
cable
- High
Voltage cable
- other
cables
- Check
that the HV shutdown (inhibit) is plugged from the preamp power supply to
the HV supply.
- Make
sure HV supply is set to zero then turn on NIM bin/Preamp power supply.
- Check
LED on the detector, it has three conditions:
- Off:
Detector is cold and ready to accept HV.
- Continuous
On: Detector is too warm
Ø
If it was cold, one must wait until detector is warm
before cooling again
Ø
If it was not yet cold, one must continue to fill
detector and wait longer
Ø
Blinking: Detector is at room temperature at it
is safe to cool down again
Ø
If case 1 is met above: recorded the peak-to-peak
noise on the central contact preamp, preamp currents, etc.
- Before
applying power check that there is noise on all the segment preamps.
If there is no signal from one of the preamps, do not apply high voltage!
- Check
that the HV setting is Positive and the Shutdown signal is Ortec type.
- While
monitoring the preamp signal with the oscilloscope, slowly increase the
voltage on the detector. The oscilloscope has to be setup on 200 mV energy
scales and ~50us time scales. The preamp signal will initially “move
up” but later recover. Do not allow the baseline to jump more that
200 mV when you increase the HV. When it recovers it is safe to apply more
voltage.
- Verify
the maximum operating voltage for the detector from the individual
specification sheets and check if it will hold this bias.